| Sign In | Join Free | My benadorassociates.com |
|
Brand Name : KACISE
Model Number : KGD-YQ-534
Certification : CE
Place of Origin : CHINA
MOQ : 1PCS
Payment Terms : L/C,D/A,D/P,T/T,Western Union,MoneyGram
Supply Ability : 1000 Piece/Pieces per Week negotiable
Delivery Time : 5-8 work days
Packaging Details : each unit has individual box and all boxes are packed in standard packages or customers requests available
Response Time : < 30s
Working Temperature : -10 - 50°C
Working Humidity : 0 - 98%RH (Non - condensing)
Working Voltage : 12V ± 10%
Output Method : RS - 485/RS - 232
Storage Conditions : -40 - 85°C; 0 - 98%RH (Non - condensing)
Laser Sensor Module KGD - YQ - 534
Product Introduction
The KGD - YQ - 534 Laser Sensor Module is a high - performance gas analysis sensor module. Based on tunable laser absorption spectroscopy technology, it uses a high - performance probe and a gas cell processed by a special process, featuring high accuracy, high sensitivity, strong anti - interference ability, long service life, and stable performance.
The product adopts a modular design for easy installation. Operations such as calibration and calibration can be carried out on the product through RS - 232/485 communication interfaces. The constant - temperature design effectively avoids the influence of temperature changes on the sensor, enabling it to work stably in harsh working conditions and complex gas environments.

Product Features
Technical Parameters
| Measurement Principle | Tunable Diode Laser Absorption Spectroscopy (TDLAS) | |||||||||||
| Measured Gas | O₂ | O₂ | O₂ | O₂ | CO | CO | CO₂ | H₂S | CH₄ | CH₄ | C₂H₂ | HF |
| Range | 0 - 10000ppm | 0 - 5%VOL | 0 - 25%VOL | 0 - 100%VOL | 0 - 500ppm | 0 - 1000ppm | 0 - 1000ppm | 0 - 1000ppm | 0 - 100ppm | 0 - 1000ppm | 0 - 100ppm | 0 - 500ppm |
| Accuracy | ± 2%F.S. | ± 2%F.S. | ± 1%F.S. | ± 0.5%F.S. | ± 2%F.S. | ± 1%F.S. | ± 2%F.S. | ± 2%F.S. | ± 2%F.S. | ± 1%F.S. | ± 2%F.S. | ± 2%F.S. |
| Resolution | 1ppm | 0.01% | 0.01% | 0.01% | 0.1ppm | 0.1ppm | 0.1ppm | 0.1ppm | 0.1ppm | 0.1ppm | 0.1ppm | 0.1ppm |
| Response Time | < 30s | |||||||||||
| Working Temperature | -10 - 50°C | |||||||||||
| Working Humidity | 0 - 98%RH (Non - condensing) | |||||||||||
| Working Voltage | 12V ± 10% | |||||||||||
| Output Method | RS - 485/RS - 232 | |||||||||||
| Storage Conditions | -40 - 85°C; 0 - 98%RH (Non - condensing) | |||||||||||
| Constant Temperature | 50°C | |||||||||||
| Dimensions (LWH) | 278*90*70mm | |||||||||||
|
|
RS 485/RS 232 Output Method Laser Sensor Module KGD YQ 534 Advanced Technology for Measurement Images |